LEVEL 1? DISPOSABLE NORMOTHERMIC IV ADMINISTRATION SET D-300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2017-12-21 for LEVEL 1? DISPOSABLE NORMOTHERMIC IV ADMINISTRATION SET D-300 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[95491505] It was reported that a leak on the y-connector of a level 1? Disposable normothermic iv administration set during intravenous and blood transfusion. The incident occurred immediately upon use. It was reported that there was no patient injury and no medical intervention required.
Patient Sequence No: 1, Text Type: D, B5


[116792025] The third sample of a y-connector of a level 1? Disposable normothermic iv administration set was not returned. The complaint states this is the third of 3 sets where a leak on the y connector was observed. This device evaluation is based on the evaluation of the first 2 samples. The samples were connected to an h-1200 hotline fluid warmer for functional testing. The samples leaked at the joint of the 3 "y" connector. A review of the manufacturing and inspection processes was conducted and was considered adequate and correct. The failure mode was leaking. The root cause was identified as a lack of robust and repeatable adhesive dispensing to achieve the desired adhesive depth and that the components were potentially not tested for leaks. See mfr: 3012307300-2017-02640, 3012307300-2017-02641.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3012307300-2017-02642
MDR Report Key7137809
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2017-12-21
Date of Report2018-01-17
Date of Event2017-09-04
Date Mfgr Received2018-01-11
Date Added to Maude2017-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone7633833310
Manufacturer G1SMITHS HEALTHCARE MFG
Manufacturer StreetS.A. DE C.V. AVE CALIDAD NO. 4 PARQUE INDUSTRIAL INTERNACIONA
Manufacturer CityTIJUANA, 22425
Manufacturer CountryMX
Manufacturer Postal Code22425
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEVEL 1? DISPOSABLE NORMOTHERMIC IV ADMINISTRATION SET
Generic NameDEVICE, WARMING. BLOOD AND PLASMA
Product CodeKZL
Date Received2017-12-21
Returned To Mfg2017-09-26
Catalog NumberD-300
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.