QUART ARTERIAL FILTER WITH SOFTLINE COATING HBF 140-J 701031617

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-21 for QUART ARTERIAL FILTER WITH SOFTLINE COATING HBF 140-J 701031617 manufactured by Maquet Cardiopulmonary Ag.

Event Text Entries

[95791690] (b)(4). Maquet getinge group is aware of similar complaints from this product. Similar products, showing a similar malfunction, have been tested. We tested the returned filter in our laboratory, following our standard process, on tightness. Therefore, the quart was tested at below water in a water bath and was then pressurized with compressed air (0. 3 bar). Due to the air escaping (visible as air bubbles in water) a leakage in the welding cover has been found. The leakage at the connection between cover and filter body can be confirmed. Most possible root cause could be the bad welding between cover and filter body. Since the reported failure did not contribute to a death or serious injury no corrective action is needed. In addition at this time it cannot be concluded that this is an systemic error. This data will be handled through a designated maquet trending process. If a trend occurs, it will be escalated to quality assurance management for review and determination if further investigation is necessary.
Patient Sequence No: 1, Text Type: N, H10


[95791691] Customer reported that "during priming at (b)(6), it was noted that the priming solution was leaking from the upper side of the quart. " no known consequences to the patient. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010762-2017-00406
MDR Report Key7138656
Date Received2017-12-21
Date of Report2017-12-21
Date of Event2017-11-02
Date Mfgr Received2017-12-21
Device Manufacturer Date2016-06-01
Date Added to Maude2017-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMAQUET CARDIOPULMONARY AG KEHLER STRASSE 31
Manufacturer City76437 RASTATT
Manufacturer CountryGM
Manufacturer Phone4972229321
Manufacturer G1BERND RAKOW
Manufacturer StreetMAQUET CARDIOPULMONARY AG KEHLER STRASSE 31
Manufacturer City76437 RASTATT
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameQUART ARTERIAL FILTER WITH SOFTLINE COATING
Generic NameFILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE
Product CodeDTM
Date Received2017-12-21
Model NumberHBF 140-J
Catalog Number701031617
Lot Number70112831
Device Expiration Date2018-06-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CARDIOPULMONARY AG
Manufacturer AddressRASTATT GM


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-21

© 2024 FDA.report