AQUABPLUS 2000 G02040107-US

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-12-21 for AQUABPLUS 2000 G02040107-US manufactured by Vivonic Gmbh.

Event Text Entries

[95667349] A plant investigation is in progress. A supplemental medwatch report will be submitted upon completion of this activity.
Patient Sequence No: 1, Text Type: N, H10


[95667350] An installation logistics manager for vivonics reported that an aquabplus reverse osmosis machine underwent troubleshooting for a power failure. During machine repair, the manager had discovered that the power-switch terminal connections were melted. There was no report of heat damage to the machine? S other components. The area technical operations manager did report evidence of arching. The power-switch is being returned to the manufacturer for physical evaluation. No further information has been made available at this time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010850471-2017-00003
MDR Report Key7139551
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-12-21
Date of Report2018-03-28
Date of Event2017-12-01
Date Mfgr Received2018-03-21
Date Added to Maude2017-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTHOMAS C. JOHNSON
Manufacturer Street920 WINTER ST.
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7816999499
Manufacturer G1VIVONIC GMBH
Manufacturer StreetKURUFUERST-EPPSTEIN-RING 4
Manufacturer CitySAILAUF 63877
Manufacturer CountryGM
Manufacturer Postal Code63877
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAQUABPLUS 2000
Generic NameSUBSYSTEM, WATER PURIFICATION
Product CodeFIP
Date Received2017-12-21
Returned To Mfg2017-12-21
Model NumberAQUABPLUS 2000
Catalog NumberG02040107-US
Lot NumberN/A
Device AvailabilityR
Device AgeMO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVIVONIC GMBH
Manufacturer AddressKURUFUERST-EPPSTEIN-RING 4 SAILAUF 63877 GM 63877


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-21