UNICORE DRILL SIZE 31 REFILL 7121

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-12-21 for UNICORE DRILL SIZE 31 REFILL 7121 manufactured by Ultradent Products Inc..

Event Text Entries

[95852639] Though no medical/surgical intervention was required to preclude a serious injury in this event, there have been previously reported events involving a similar device that resulted in the need for medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function. Therefore, this event meets the criteria for reportability per 21 cfr part 803. After evaluation of returned drill it was determined that the drill had previously been damaged by being grasped tightly on the cutting edge by a device which may have been used to insert the drill into the hand piece. Refile due to issues with uploading to maude data base
Patient Sequence No: 1, Text Type: N, H10


[95852640] Dentist had a #1 unicore drill break in the canal he was able to retrieve the broken portion of the drill from the tooth and proceed with procedure. Dentist reported all is fine with the tooth.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1718912-2017-00040
MDR Report Key7139675
Date Received2017-12-21
Date of Report2017-12-21
Date of Event2015-02-11
Date Mfgr Received2015-02-11
Device Manufacturer Date2013-09-05
Date Added to Maude2017-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS RONA MURPHY
Manufacturer Street505 WEST 10200 SOUTH
Manufacturer CitySOUTH JORDAN UT 84095
Manufacturer CountryUS
Manufacturer Postal84095
Manufacturer Phone8005525512
Manufacturer G1ULTRADENT PRODUCTS
Manufacturer Street505 WEST 10200 SOUTH
Manufacturer CitySOUTH JORDAN UT 84095
Manufacturer CountryUS
Manufacturer Postal Code84095
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameUNICORE DRILL SIZE 31 REFILL
Generic NameUNICORE DRILL
Product CodeEJL
Date Received2017-12-21
Returned To Mfg2015-02-23
Catalog Number7121
Lot NumberB8NYL
OperatorDENTIST
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerULTRADENT PRODUCTS INC.
Manufacturer Address505 WEST 10200 SOUTH SOUTH JORDAN UT 84095 US 84095


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-12-21

© 2026 FDA.report