[95760783] Beginning in (b)(6) 2017 through (b)(6) 2017, there were 5 instances where a set of 5 samples appeared to be missing, but were found on the roche mpa. In working with the vendor it was eventually determined that since the lab has 2 mpa instruments that racks from mpa 1 were being used on mpa2. These racks are identical including the bar code numbers with the exception of a green sticker for mpa. When the second rack with the same bar code number is introduced to the analyzer and the first rack with the same number is currently being processed, the analyzer will produce an error message that indicates "rack and pos is already occupied by another sample. " the instrument allows the operator to override the error and re-introduce the rack. The instrument than disassociates the first set of pt demographics and replaces it with the second set of pt demographics, thereby testing the 1st set of samples and applying the second set of pt demographics to the result. This cause 22 results to be reported on the incorrect pts.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5074195
MDR Report Key7142245
Date Received2017-12-21
Date of Report2017-12-20
Date Added to Maude2017-12-22
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameROCHE
Generic NameMPA (MYCOPHENOLIC ACID ASSAY)
Product CodeDLJ
Date Received2017-12-21
Model NumberMPA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerROCHE
Manufacturer AddressBASEL, CH


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-12-21