COBAS 8000 E 801 MODULE E801

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-12-22 for COBAS 8000 E 801 MODULE E801 manufactured by Roche Diagnostics.

Event Text Entries

[95822841] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[95822842] The customer received a questionable low elecsys ca 125 ii result for one patient sample. The initial result was 0. 633 u/ml. The repeat results on (b)(6) 2017 were 59 and 58. 2 u/ml. No erroneous result was reported outside of the laboratory. There was no allegation of an adverse event. The reagent lot number 286641. The expiration date was requested but was not provided. The calibration and qc results on the day of the event were in range.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-03129
MDR Report Key7142473
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-12-22
Date of Report2018-01-11
Date of Event2017-12-13
Date Mfgr Received2017-12-15
Date Added to Maude2017-12-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 8000 E 801 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeLTK
Date Received2017-12-22
Model NumberE801
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 8000 E 801 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-12-22
Model NumberE801
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.