MERGE EYE STATION MERGE EYE STATION V11.5

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-12-29 for MERGE EYE STATION MERGE EYE STATION V11.5 manufactured by Merge Healthcare.

Event Text Entries

[96235223] Technical support found the issue was related to the small form factor. This report is related to recall 2183926-08/19/2016-073-c/z-1142-2017. The resolution of switching the driver back to the legacy driver is the resolution. According to information received from the customer, switching back to the legacy driver seems to have resolved the issue.
Patient Sequence No: 1, Text Type: N, H10


[96235224] Merge eye station is intended to be used in conjunction with existing ophthalmic fundus caeras to take images of the eye, perform fluorescein angiography, red free, color and icg still-image photography as well as video imaging. On 24oct2017, merge healthcare was notified that camera was freezing mid procedure. The system was switched back to the legacy driver, as outlined in recall z-1142-2017. It was originally thought the complaint has been called in previously in another case ((b)(4)). Follow-up occurred with the account and on 04dec2017, the account reported that this issue was actually a separate issue for another location. The account reported that before the legacy driver switch, the issue was occurring randomly and unexpectedly and at numerous times. According to the customer, the issue led in rescheduling of patients for fa (fluorosceine angiogram) exams. To date, the issue has not occurred again. This issue is being reported due to the potential for harm related to the inability of the eye care professional to obtain the necessary information for procedures, in a timely manner. No patient harm occurred as a result of this issue. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183926-2017-00209
MDR Report Key7156198
Report SourceHEALTH PROFESSIONAL
Date Received2017-12-29
Date of Report2017-12-04
Date of Event2017-10-24
Date Mfgr Received2017-12-04
Device Manufacturer Date2014-12-15
Date Added to Maude2017-12-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MEG MUCHA
Manufacturer Street900 WALNUT RIDGE DR.
Manufacturer CityHARTLAND, WI 53029
Manufacturer CountryUS
Manufacturer Postal53029
Manufacturer Phone2629123514
Manufacturer G1MERGE HEALTHCARE
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND, WI 53029
Manufacturer CountryUS
Manufacturer Postal Code53029
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-1142-2017
Event Type3
Type of Report3

Device Details

Brand NameMERGE EYE STATION
Generic NameCAMERA, OPHTHALMIC, AC-POWERED
Product CodeHKI
Date Received2017-12-29
Model NumberMERGE EYE STATION V11.5
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMERGE HEALTHCARE
Manufacturer Address900 WALNUT RIDGE HARTLAND, WI 53029 US 53029


Patients

Patient NumberTreatmentOutcomeDate
10 2017-12-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.