XEMPLIFI DBM PUTTY 8103.0010S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-01-02 for XEMPLIFI DBM PUTTY 8103.0010S manufactured by Allosource.

Event Text Entries

[96240522] According to the report from the distributor, an infection had materialized 6 weeks post surgery. The tissue based medical device is aseptically processed, freeze dried, and terminally sterilized via electron beam irradiation. 77 total tissue lots were produced from donor (b)(6): 25 device lots, 45 bone grafts, 7 soft tissue grafts. 67 total lots were distributed: 21 device lots were distributed, 4 used as qc samples; 39 bone lots were distributed, 6 were discarded prior to distribution; all 7 soft tissue lots were distributed. One lot of tissue based device was implanted into the patient associated with this adverse event (b)(4). Allosource has received no additional reports of adverse reactions or adverse events associated with the other 66 distributed lots. 45 total tissue lots were produced from donor (b)(6): 25 device lots, 20 bone grafts. 41 total lots were distributed: 20 device lots, 4 used as qc samples, 1 discarded prior to distribution; 16 bone lots were distributed, and 4 were quarantined. Three lots of tissue based device were implanted into the patient associated with this adverse event (b)(4). Allosource has received no additional reports of adverse reactions or adverse events associated with the other 38 distributed lots. This report was not submitted on the date prepared for submission due to technical issues with esg. These are the first emdrs to be reported electronically by allosource since initiation of the system, and the individual who had the esg account for submission is no longer with the company. Additionally, the original certificate required to initiate the webtrader account with esg was on a device that was decommissioned and it could not reproduce it. These technical issues were discussed with an individual with the emdr group at cdrh who informed the reporter to place a note in section explaining the technical issues.
Patient Sequence No: 1, Text Type: N, H10


[96240523] Four lots of xemplifi dbm putty 10cc, from two donors, manufactured by allosource were implanted into a patient on (b)(6) 2017. On (b)(6) 2017 the surgeon notified the private label distributor that he was worried that the patient got a staph infection from the "bone graft". Allosource received notification of the incident on (b)(6) 2017. The four lots of tissue based medical devices remain in the patient. No additional information has been received from the surgeon to date. Four separate adverse event reports are being submitted for the event associated with this patient: 3000215346-2017-00009, 3000215346-2017-00010, 3000215346-2017-00011, 3000215346-2017-00012.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3000215346-2017-00010
MDR Report Key7158038
Report SourceDISTRIBUTOR
Date Received2018-01-02
Date of Report2017-09-28
Date of Event2017-08-17
Date Mfgr Received2017-08-30
Device Manufacturer Date2017-03-21
Date Added to Maude2018-01-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TREVOR WRIGHT
Manufacturer Street6278 S. TROY CIRCLE
Manufacturer CityCENTENNIAL CO 80111
Manufacturer CountryUS
Manufacturer Postal80111
Manufacturer Phone7208734733
Manufacturer G1ALLOSOURCE
Manufacturer Street6278 S. TROY CIRCLE
Manufacturer CityCENTENNIAL CO 80111
Manufacturer CountryUS
Manufacturer Postal Code80111
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXEMPLIFI DBM PUTTY
Generic NameBONE VOID FILLLER
Product CodeMBP
Date Received2018-01-02
Model Number8103.0010S
Device Expiration Date2019-03-21
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALLOSOURCE
Manufacturer Address6278 S. TROY CIRCLE CENTENNIAL CO 80111 US 80111


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-01-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.