EDGEPLUS VALVED ENTRY SYSTEM 8065751657

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-03 for EDGEPLUS VALVED ENTRY SYSTEM 8065751657 manufactured by Alcon Research, Ltd. - Irvine Technology Center.

Event Text Entries

[96394732] No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted. No sample has been returned for evaluation; therefore, the condition of the product could not be verified. Due to an observed increased trend for this event, an internal investigation was initiated to determine if corrective and preventative actions were necessary. A root cause for the increased complaint rate was found to be related to a manufacturing post assembly inspection practice. This complaint does not identify a reported lot and no sample was returned for evaluation so it cannot be determined if the complaint can be attributed to the post assembly inspection. The post assembly inspection has been discontinued and an effectiveness check has been established and will be monitored periodically. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[96394734] A customer reported experiencing ongoing issues with leaking trocars. Patient outcome is unknown. The reporter indicated that no further information was available. No product samples are available for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2028159-2018-00016
MDR Report Key7159700
Date Received2018-01-03
Date of Report2018-01-03
Date Mfgr Received2017-12-06
Date Added to Maude2018-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NADIA BAILEY
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8176152230
Manufacturer G1ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Street15800 ALTON PARKWAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameEDGEPLUS VALVED ENTRY SYSTEM
Generic NameCANNULA, TROCAR, OPHTHALMIC
Product CodeNGY
Date Received2018-01-03
Model NumberNA
Catalog Number8065751657
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Address15800 ALTON PARKWAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-01-03