CHROMID? S. AUREUS ELITE 419042

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-01-03 for CHROMID? S. AUREUS ELITE 419042 manufactured by Biom?rieux Espa?a.

Event Text Entries

[97411749] A customer from (b)(6) reported to biom? Rieux a false negative result for a (b)(6) strain on a chromid? S. Aureus elite (lot 1005950070) plate. The customer stated a specific strain of staphylococcus aureus did not grow on the chromid? S. Aureus elite plate. The strain was cultured on chromid? S. Aureus elite, chromid? (b)(6) smart, and blood agar. The s. Aureus only grew on the blood agar. The customer then performed disc testing and found the strain to be cefoxitin resistant. It was then cultured from the original sample tube to a new s. Aureus elite plate and a (b)(6) smart plate, and there was no growth on either plate. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00003
MDR Report Key7162306
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-01-03
Date of Report2018-04-05
Date Mfgr Received2018-03-13
Device Manufacturer Date2017-09-07
Date Added to Maude2018-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer StreetISAAC NEWTON, 6 TRES CANTOS
Manufacturer CityMADRID, SPAIN 28760
Manufacturer CountrySP
Manufacturer Postal Code28760
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? S. AUREUS ELITE
Generic NameCHROMID? S. AUREUS ELITE
Product CodeJSI
Date Received2018-01-03
Catalog Number419042
Lot Number1005950070
ID Number03573026494445
Device Expiration Date2017-12-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX ESPA?A
Manufacturer AddressISAAC NEWTON, 6 TRES CANTOS MADRID, SPAIN 28760 SP 28760


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-03

© 2024 FDA.report