COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER 6605526

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2006-05-17 for COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER 6605526 manufactured by Beckman Coulter, Inc..

Event Text Entries

[448293] A customer contacted beckman coulter regarding erroneous rbc, hemoglobin (hgb), and hematocrit (hct) results generated by the coulter hmx hematology analyzer with autolader. Per customer, on 04/07/06 a cross-match specimen was collected from a patient and tested for cbc. Rbc, hgb and hct results were higher when compared with results obtained a day earlier; the customer re-tested the patient sample and the rebc hgb and hct results were in the same range as in the 1st run of 04/07/06; an hour later, the customer collected a fresh sample for the patient and tested for cbc; rhe rbc, hgb and hct results were totally different, but closely mathced results from 1 day before; the sample was re-tested and rbc, hgb, and hct repeated results mathced the results from event date. For the actual rbc, hgb, and hct results please see b6. Baed on available information, there was no affect to patient treatment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1061932-2006-00017
MDR Report Key716590
Report Source05
Date Received2006-05-17
Date of Report2006-05-17
Date of Event2006-04-07
Date Mfgr Received2006-04-18
Device Manufacturer Date2001-05-01
Date Added to Maude2006-05-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactNORA ZERUNIAN, MGR
Manufacturer Street200 S. KRAEMER BLVD P.O BOX 8000 MAIL STOP: W-110
Manufacturer CityBREA CA 928228000
Manufacturer CountryUS
Manufacturer Postal928228000
Manufacturer Phone7149613634
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Generic NameHEMATOLOGY ANALYZER
Product CodeGKH
Date Received2006-05-17
Model NumberAUTOLOADER
Catalog Number6605526
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key705658
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address11800 SW 147TH AVE. MIAMI FL 33196 US
Baseline Brand NameCOULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Baseline Generic NameHEMATOLOGY ANALYZER
Baseline Model NoAUTOLOADER
Baseline Catalog No6605526
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-05-17

© 2026 FDA.report