STEAM TAPE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-04 for STEAM TAPE manufactured by Steris Biological Operations.

Event Text Entries

[97498092] No report of procedure delay or cancellation. The instruments present during the time of the reported event were reprocessed before use. The user facility obtained the verify steam tape subject of the reported event from a steris distributor, owens & minor. Following the reported event, user facility personnel contacted owens & minor to describe the event that took place. The verify steam tape instructions for use states (pg. 1), "expose pack to the steam sterilization process. Following processing, remove the pack and examine the tape. The white indicator bars will be brown to black in color if exposed to steam sterilization conditions. " retain testing was performed on the lot number subject of the reported event; no issues were noted. The dhr for the subject lot was reviewed and no abnormalities were found. The user facility is returning on hand inventory for the lot subject of the reported event to steris. Steris will offer in-service training on the proper use of the verify steam tape to the user facility. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[97498093] The user facility reported via medwatch report #(b)(4) that their verify steam tape did not change colors following a completed cycle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004080920-2018-00001
MDR Report Key7166028
Date Received2018-01-04
Date of Report2018-01-04
Date Mfgr Received2017-12-05
Date Added to Maude2018-01-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927453
Manufacturer G1STERIS BIOLOGICAL OPERATIONS
Manufacturer Street9325 PINECONE DR.
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal Code44060
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTEAM TAPE
Generic NameSTEAM TAPE GENERAL PURPOSE STEAM INDICATOR TAPE
Product CodeJOJ
Date Received2018-01-04
Lot Number709401
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS BIOLOGICAL OPERATIONS
Manufacturer Address9325 PINECONE DR MENTOR OH 44060 US 44060


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-04