STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-01-05 for STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[96617306] Patient's exact age is unknown; however it was reported that the patient was over the age of 18. (b)(4). Although the suspect device has been received, the evaluation has not been completed. Therefore, the cause of the reported malfunction has not been determined. Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[96617307] It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was to be used in the ureter during a ureteroscopy (urs) procedure performed on (b)(6) 2017. According to the complainant, during unpacking, a hole was noted in the sterile packaging. The procedure was completed with another stone cone nitinol urological retrieval coil. The patient's condition at the conclusion of the procedure was reported to be stable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-00013
MDR Report Key7169415
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-01-05
Date of Report2017-12-19
Date of Event2017-10-26
Date Mfgr Received2018-01-25
Device Manufacturer Date2017-04-03
Date Added to Maude2018-01-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2018-01-05
Returned To Mfg2018-01-02
Model NumberM0063903200
Catalog Number390-320
Lot NumberG35698
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.