LIGHT SOURCE, 500XL, XENON 72200568

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2018-01-08 for LIGHT SOURCE, 500XL, XENON 72200568 manufactured by Smith & Nephew, Inc..

Event Text Entries

[96612515]
Patient Sequence No: 1, Text Type: N, H10


[96612516] It was reported that the light was very weak. There was a delay over 60 minutes, there was no patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


[130373037] The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed. A relationship, if any, between the subject device and the reported event could not be determined.
Patient Sequence No: 1, Text Type: N, H10


[135384823] The device was received for evaluation. There was a relationship found between the returned device and the reported incident. A visual inspection was performed on the product and no issue was observed. Complaint of low light output was confirmed. The lumen output minimum spec is 1600 lm and unit? S output was 755 lm. Cause of low light output is a cracked glass rod, a part of the clad rod assembly. Except for the glass rod no other damage was observed on the exterior or interior of unit. Cause of damage to glass rod could not be determined. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1643264-2018-00026
MDR Report Key7171448
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2018-01-08
Date of Report2018-04-03
Date of Event2017-12-16
Date Mfgr Received2018-03-26
Date Added to Maude2018-01-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street7000 WEST WILLIAM CANNON DRIVE
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer G1SMITH & NEPHEW, INC.
Manufacturer Street76 S. MERIDIAN AVE.
Manufacturer CityOKLAHOMA CITY OK 731076512
Manufacturer CountryUS
Manufacturer Postal Code731076512
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIGHT SOURCE, 500XL, XENON
Generic NameIMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Product CodeFFS
Date Received2018-01-08
Returned To Mfg2018-03-12
Model Number72200568
Catalog Number72200568
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address76 S. MERIDIAN AVE. OKLAHOMA CITY OK 731076512 US 731076512