P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER 10065

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-01-09 for P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER 10065 manufactured by Vyaire Medical, Inc.

Event Text Entries

[96773718] Vyaire file identification: (b)(4). Any additional information received from the customer will be included in a follow-up report. Results of investigation: the vyaire failure analysis laboratory received the suspect microblender assembly for investigation. An evaluation of the component could be confirmed and duplicated. The dry o-rings on the poppet assembly was causing the poppet to stick and not alarm. The alarm adjustments were also found to be out of calibration.
Patient Sequence No: 1, Text Type: N, H10


[96773719] The customer reported while using the p/n 9320 low flow air/oxygen microblender; the device's alarm is not working. The customer reported the device is due for overhaul and repairs. The customer reported there is no patient involvement associated with the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2018-07271
MDR Report Key7175631
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-01-09
Date of Report2018-01-09
Date of Event2017-12-20
Date Mfgr Received2017-12-20
Device Manufacturer Date2008-08-01
Date Added to Maude2018-01-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN GRAF
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer G1VYAIRE MEDICAL, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameP/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Generic NameMIXER, BREATHING GASES, ANESTHESIA INHALATION
Product CodeBZR
Date Received2018-01-09
Returned To Mfg2017-12-28
Model NumberMICROBLENDER
Catalog Number10065
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-09

© 2024 FDA.report