NEOBLUE 001103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-01-09 for NEOBLUE 001103 manufactured by Natus Medical Incorporated.

Event Text Entries

[96823536] Replacement part has been sent to the customer to resolve the issue. Additional information has been requested from customer, for further investigation. No further investigation possible, if additional information becomes available natus will provide follow-up report.
Patient Sequence No: 1, Text Type: N, H10


[96823537] On (b)(6) 2017, the customer reported to natus about their neoblue 3 led light output intensity measured low using natus nb radiometer. The customer stated that maximum output with the neoblue radiometer read 25-27 [? ]w/cm2/nm when measured with the natus radiometer at 12 inches below the center. The customer was unable to adjust the potentiometer up and down. The last time the nb device measured by biomed was in (b)(6) 2017, and the output was 30. 5 [? ]w/cm2/nm. The customer confirmed there was delay in treatment, but no death/serious injury or environmental/safety concerns.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00022
MDR Report Key7177063
Report SourceUSER FACILITY
Date Received2018-01-09
Date of Report2017-12-12
Date Mfgr Received2017-12-12
Device Manufacturer Date2017-07-07
Date Added to Maude2018-01-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARC RIVAS
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685142
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEOBLUE
Generic NameNEOBLUE 3
Product CodeLBI
Date Received2018-01-09
Model Number001103
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-09