ENDOSONIC CLEANER 100V (OT) 7500617

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-10 for ENDOSONIC CLEANER 100V (OT) 7500617 manufactured by Keymed (medical And Industrial Equipment) Ltd..

Event Text Entries

[97480954] Okm m-bc will be requesting the product to be returned for further investigation. No alleged report of injury to patient or user.
Patient Sequence No: 1, Text Type: N, H10


[97480955] Endosonic arrived with nagano olumpus at the request for repair, because there was a description that seemed to have been shocked to the content of the request, it was confirmed to the dealer that there was no health damage but the nurse was turned on (ultrasound off) when she touched the tank with a state, she received a report that she received a feeling that the electricity flowed like flowing. This device confirmed the leakage at nagano ol, but there was no abnormality. Repair request content: the service got a repair request the other day and it was returned without any abnormality and it seemed that there was electric leak when checking at the facility. Put a switch for checking the unrepaired item, and when the nurse touched the cleaning tank, the service felt like the electricity flowed. Please check again. Last repair request (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611174-2018-00002
MDR Report Key7177740
Date Received2018-01-10
Date of Report2018-04-16
Date Mfgr Received2018-04-04
Device Manufacturer Date2010-12-31
Date Added to Maude2018-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR PETER MORCOS
Manufacturer StreetKEYMED HOUSE, STOCK ROAD
Manufacturer CitySOUTHEND ON SEA, ESSEX
Manufacturer CountryUS
Manufacturer Phone0441702616
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameENDOSONIC CLEANER 100V (OT)
Generic NameENDOSONIC CLEANER
Product CodeFLG
Date Received2018-01-10
Returned To Mfg2017-12-25
Model Number7500617
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Manufacturer AddressKEYMED HOUSE, STOCK ROAD SOUTHEND-ON-SEA, ESSEX US


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.