OLYMPIC COOL-CAP 401352-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-01-10 for OLYMPIC COOL-CAP 401352-01 manufactured by Natus Medical Inc..

Event Text Entries

[96966653] The customer reported that the patient was placed into a backup cooling device. The customer has not reported the alarm identifier, the patient's rectal temperature or the specific hours of treatment. Follow up communication attempts with the customer were carried out on 14 dec 2017 and 09 jan 2018. These follow up attempts have been unsuccessful in obtaining additional information thus far. The customer has since placed an order for device service and a loaner cool cap unit.
Patient Sequence No: 1, Text Type: N, H10


[96966654] A customer called natus technical service to report that, prior to the call, the cool cap device was indicating that there was a major water leak and the patient had been cooling for two days. The customer confirmed that there was no apparent water leakage, but the main hose (red) was reported not to be completely connected. The customer also reported a full water bag. After properly connecting the main hose, the issue still persisted, and the customer was unable to clear the error. The cool cap was rebooted, and during the reboot process, the customer reported that the system would not fill and that the resume selection was "grayed out. " the customer attempted to replace/spike a new bag. The customer also reported that the area was loud, which would not allow them to verify if the pump was active or if the system made any clicking noises. No death, serious injury, delay in care or environmental/safety concerns have been reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00030
MDR Report Key7179627
Report SourceUSER FACILITY
Date Received2018-01-10
Date of Report2017-12-13
Date Mfgr Received2018-02-07
Device Manufacturer Date2008-09-30
Date Added to Maude2018-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2018-01-10
Returned To Mfg2018-02-07
Model Number401352-01
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC.
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-01-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.