COBAS E 411 IMMUNOASSAY ANALYZER E411 DISK 4775279001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-01-10 for COBAS E 411 IMMUNOASSAY ANALYZER E411 DISK 4775279001 manufactured by Roche Diagnostics.

Event Text Entries

[97737260] (b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[97737261] The customer received a discrepant (b)(6) elecsys anti-hcv ii immunoassay result for a patient sample on the cobas e 411 immunoassay analyzer. The initial result was (b)(6) and was reported outside of the laboratory. The customer repeated the sample because the patient had an (b)(6) result of (b)(6) on (b)(6) 2017. The repeat result was (b)(6). The calibration and quality control were acceptable. There was no allegation of an adverse event. The (b)(6) reagent lot number was 26719301. The expiration date was requested but not provided.
Patient Sequence No: 1, Text Type: D, B5


[131710866] The patient was born in (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[131710969] The sample was investigated and determined to be positive. The customer's initial negative result could not be reproduced. The investigation was unable to find a definitive root cause.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2018-00113
MDR Report Key7179734
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-01-10
Date of Report2018-03-19
Date of Event2017-12-24
Date Mfgr Received2017-12-27
Date Added to Maude2018-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeMZP
Date Received2018-01-10
Model NumberE411 DISK
Catalog Number4775279001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2018-01-10
Model NumberE411 DISK
Catalog Number4775279001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.