IMRIS T2X OR TABLE V2 (ORT100) 110470-000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-12 for IMRIS T2X OR TABLE V2 (ORT100) 110470-000 manufactured by Imris - Deerfield Imaging, Inc..

Event Text Entries

[97876641] This table was manufactured in 2010 and had been in use for 8 years. An internal capa ((b)(4)) was initiated for drift observed in ort100 tables. The table in question was repaired using corrective and preventive actions implemented in the capa. The hydraulic fluid was flushed and the in-line filter was replaced. After the hydraulic fluid exchange was performed, an air bleed procedure was conducted. The table was then drift tested to verify proper function.
Patient Sequence No: 1, Text Type: N, H10


[97876642] During a patient procedure, it was observed that the ort100 operating room table was drifting slightly to the right. The level of drift was observed via the catheter bag not being level. Or personnel re-leveled the table approximately every half hour during the procedure and were able to complete the procedure successfully. There were no injuries sustained by the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010326005-2018-00001
MDR Report Key7188170
Date Received2018-01-12
Date of Report2018-01-12
Date of Event2017-12-21
Date Mfgr Received2017-12-21
Device Manufacturer Date2010-02-15
Date Added to Maude2018-01-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD SPERLING
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal553434100
Manufacturer Phone7632036344
Manufacturer G1IMRIS - DEERFIELD IMAGING, INC.
Manufacturer Street5101 SHADY OAK ROAD
Manufacturer CityMINNETONKA MN 553434100
Manufacturer CountryUS
Manufacturer Postal Code553434100
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameIMRIS T2X OR TABLE V2 (ORT100)
Generic NameOPERATING ROOM TABLE
Product CodeKXJ
Date Received2018-01-12
Model NumberORT100
Catalog Number110470-000
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMRIS - DEERFIELD IMAGING, INC.
Manufacturer Address5101 SHADY OAK ROAD MINNETONKA MN 553434100 US 553434100


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.