DRILL, FLEXIBLE, ENDOSCPC, CANN, 4.5MM 72202971

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-15 for DRILL, FLEXIBLE, ENDOSCPC, CANN, 4.5MM 72202971 manufactured by Smith & Nephew, Inc..

Event Text Entries

[97489708]
Patient Sequence No: 1, Text Type: N, H10


[97489709] It was reported that the tip of the drill broke during the perforation of the bone tunnel. All parts were removed from the patient, there was no damage to the patient.
Patient Sequence No: 1, Text Type: D, B5


[106885033] One 4. 5mm flexible endoscopic drill was returned for evaluation. Visual assessment of the drill confirmed the reported breakage. The entire drill head has broken off where it transitions to the double helix of the shaft. There are no material voids at the break area. The break area is slightly uncoiled. Examination of the drill head showed it is heavily burred. Dimensional and material condition was inspected and found to meet print specification. Per the device ifu under precautions? Use of drills that have worn or dull tips can result in breakage?. No root cause related to the manufacturing process can be established. Use error may have been a contributing factor to the event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1219602-2018-00072
MDR Report Key7189128
Date Received2018-01-15
Date of Report2018-02-02
Date of Event2017-12-21
Date Mfgr Received2018-01-24
Device Manufacturer Date2015-02-18
Date Added to Maude2018-01-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street7000 WEST WILLIAM CANNON DRIVE
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer G1SMITH & NEPHEW, INC.
Manufacturer Street130 FORBES BOULEVARD
Manufacturer CityMANSFIELD MA 02048
Manufacturer CountryUS
Manufacturer Postal Code02048
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDRILL, FLEXIBLE, ENDOSCPC, CANN, 4.5MM
Generic NamePOLYMER, EAR, NOSE AND THROAT, SYNTHETIC, ABSORBABLE
Product CodeNHB
Date Received2018-01-15
Returned To Mfg2018-01-19
Catalog Number72202971
Lot Number50534270
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address130 FORBES BOULEVARD MANSFIELD MA 02048 US 02048


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-15