HUDSON ONE WAY VALVE 1644

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-01-15 for HUDSON ONE WAY VALVE 1644 manufactured by Teleflex Medical.

Event Text Entries

[97545773] Qn#(b)(4). A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device was not returned at the time of this report. The device history record (dhr) of batch number 74g1700488 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint. The dhr shows that the product was assembled and inspected according to our specifications. Customer complaint cannot be confirmed based only on the information provided. In order to perform a proper and thorough investigation to confirm the alleged defect, determine the root cause and establish any necessary corrective actions, it is necessary to evaluate the sample involved. If the device sample becomes available at a later date this report will be updated.
Patient Sequence No: 1, Text Type: N, H10


[97545774] Customer complaint alleges the device broke during use on a patient. There was no report of patient harm or consequence. There was no report of necessary medical intervention.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004365956-2018-00020
MDR Report Key7190407
Report SourceUSER FACILITY
Date Received2018-01-15
Date of Report2018-01-02
Date of Event2018-01-02
Date Mfgr Received2018-02-08
Device Manufacturer Date2017-07-04
Date Added to Maude2018-01-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON ONE WAY VALVE
Generic NameCONNECTOR, AIRWAY (EXTENSION)
Product CodeBZA
Date Received2018-01-15
Returned To Mfg2018-01-23
Catalog Number1644
Lot Number74G1700488
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-15

© 2024 FDA.report