ORS-320 SLUSH WARMER DISC DRAPE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-01-16 for ORS-320 SLUSH WARMER DISC DRAPE manufactured by Microtek Dominicana, S.a..

Event Text Entries

[97879224] A sample was received for evaluation. The drape had a small hole 22 inches from the slush plate and 7 inches to the left. Not where the drape would sit in the basin. The dhr was reviewed and it was noticed that this lot had 2400 pcs that were manufactured on 05/02/2017. No defects related to the nonconformance were reported during quality inspections. The personnel were trained. The production line this product was manufactured on has since been discontinued. The current production line was reviewed and nothing was found that could create a hole in the drape. Additionally, the method at the current production line was reviewed and it was confirmed that the operators follow the process. It was concluded that the sample sent by the customer confirmed the original complaint/description of the problem. In the sample a pinch was noticed at the hole. This pinch rises up from the bottom part of the drape and this indicates that the edges of the oven (warmer) were sharp. The location of the hole is in the same area.
Patient Sequence No: 1, Text Type: N, H10


[97879225] Cleaned and decontaminated the drape so that it can be handled. When we draped the machine the hole was near the corner of the warmer. However, there was a liter of fluid that leaked into the warmer basin. The edges of the warmer are sharp and she plans to put some duct tape on the corners to smooth them out until replacements are received. We aren't sure that is what happened but one possibility. No patient injury or treatment was reported for the incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043817-2018-00002
MDR Report Key7197080
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-01-16
Date of Report2017-12-18
Date of Event2017-12-15
Date Mfgr Received2017-12-18
Date Added to Maude2018-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS JENNIFER MRKVICKA
Manufacturer Street1 ECOLAB PLACE
Manufacturer CityST. PAUL MN 55102
Manufacturer CountryUS
Manufacturer Postal55102
Manufacturer Phone6512502237
Manufacturer G1MICROTEK DOMINICANA
Manufacturer StreetZONA FRANCA NO. 2
Manufacturer CityLA ROMANA,
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameORS-320 SLUSH WARMER DISC DRAPE
Generic NameFLUID WARMING AND SLUSH DRAPES
Product CodeLHC
Date Received2018-01-16
Returned To Mfg2018-01-02
Model NumberORS-320
Lot NumberD171211
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK DOMINICANA, S.A.
Manufacturer AddressZONA FRANCA NO. 2 LA ROMANA, DR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-16

© 2026 FDA.report