ESTEEM 2001 902001-002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-01-16 for ESTEEM 2001 902001-002 manufactured by Envoy Medical.

Event Text Entries

[97971540] Device evaluation summary: test results from device manufacturing were reviewed. Sound processor (sn (b)(4)) passed all full-functional testing. A dhr review revealed that the device met all specifications and there were no manufacturing issues. The sound processor (sn (b)(4)) was evaluated upon receipt at envoy medical. A visual inspection revealed no device abnormalities except for some surface scuffing. The battery's voltage was tested and confirmed to be depleted. The battery was confirmed to be depleted at the time of explant, though no root cause of the early battery depletion has been identified. This issue was discovered during an mdr remediation project. This report was initially submitted to the https://esgtest. Fda. Gov/ environment on 10/12/2016 rather than the https://esg. Fda. Gov/ environment. Submissions to the https://esgtest. Fda. Gov/ environment were conducted in good faith that envoy medical was following all mdr submission requirements per the 21cfr803 regulation. The issue is under corrective action investigation and this submission is part of the containment activity for that issue.
Patient Sequence No: 1, Text Type: N, H10


[97971541] Envoy medical was notified of a sound processor replacement due to battery depletion. This particular sound processor was implanted for 2. 43 years, which is less than the expected worst case battery life for normal operation of 2. 8 years. Clinical history: (b)(6) 2010 - original right ear implant by dr. (b)(6). Fittings on (b)(6) 2010, (b)(6) 2011. (b)(6) 2012 - battery depletion discovered during fitting (888 days = 2. 43 years). (b)(6) 2012 - early battery change. A visual inspection of the returned device revealed some tool scratches/gouges on the can surface. The header ports and springs are normal. These observations are consistent with rough handling during device removal. A dhr review revealed that the device met all specifications and there were no related manufacturing issues. Functional testing confirmed the battery depletion. The returned device was not interrogated to reveal its original settings when it was received in 2012. This information can no longer be obtained.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004007782-2015-00004
MDR Report Key7197775
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-01-16
Date of Report2016-10-12
Date of Event2012-07-09
Date Mfgr Received2012-08-23
Date Added to Maude2018-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD KOEPPEL
Manufacturer Street4875 WHITE BEAR PARKWAY
Manufacturer CityWHITE BEAR LAKE MN 55110
Manufacturer CountryUS
Manufacturer Postal55110
Manufacturer Phone6513618000
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameESTEEM
Generic NameESTEEM II SOUND PROCESSOR
Product CodeOAF
Date Received2018-01-16
Returned To Mfg2012-08-23
Model Number2001
Catalog Number902001-002
Lot NumberEMC0003136
Device Expiration Date2010-05-28
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerENVOY MEDICAL
Manufacturer Address4875 WHITE BEAR PARKWAY WHITE BEAR LAKE MN 55110 US 55110


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-01-16