ARCHITECT STAT CK-MB 02K42-28

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-18 for ARCHITECT STAT CK-MB 02K42-28 manufactured by Abbott Laboratories.

Event Text Entries

[98074001] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[98074002] The customer reported falsely elevated architect stat ck-mb results for two patients. The first patient (sample id (b)(6)) generated a ck-mb result of 32. 0 ng/ml. A new sample ((b)(6)) drawn 3 hours later generated a ck-mb of 2. 0 ng/ml. The second patient (sample id (b)(6)) generated a ck-mb result of 17. 9 ng/ml and retest (sid (b)(6)) 2. 3 ng/ml. A second sample drawn 3 hours later (sid (b)(6)) generated a ck-mb result of 2. 2 ng/ml and retest (sid (b)(6)) 2. 3 ng/ml. No specific patient information was provided and no adverse impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1415939-2018-00004
MDR Report Key7203536
Date Received2018-01-18
Date of Report2018-02-12
Date Mfgr Received2018-02-07
Device Manufacturer Date2017-04-01
Date Added to Maude2018-01-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600643500
Manufacturer CountryUS
Manufacturer Postal Code600643500
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameARCHITECT STAT CK-MB
Generic NameCK-MB
Product CodeJHX
Date Received2018-01-18
Catalog Number02K42-28
Lot Number79350UN17
Device Expiration Date2018-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US 600643500


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-18