RELIANCE ENDOSCOPE PROCESSOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-19 for RELIANCE ENDOSCOPE PROCESSOR manufactured by Steris Canada Corporation.

Event Text Entries

[98199594] A steris service technician arrived onsite following the reported event. While inspecting the unit, the technician noticed the braided hose connecting the facility's water supply to the unit was broken, causing the reported water leak. The technician replaced the braided hose, ran a test cycle, and confirmed the unit to be operating according to specification. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[98199595] The user facility reported water leaking from their reliance endoscope processor. No report of injury, procedure delay, or cancellations.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680353-2018-00005
MDR Report Key7206303
Date Received2018-01-19
Date of Report2018-01-19
Date of Event2017-12-21
Date Mfgr Received2017-12-21
Date Added to Maude2018-01-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927453
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC 8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameRELIANCE ENDOSCOPE PROCESSOR
Generic NamePROCESSING SYSTEM
Product CodeNZA
Date Received2018-01-19
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC 8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-19

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.