TRANSFER BELT, QUICK-RELEASE ECONOMY 6537Q

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-19 for TRANSFER BELT, QUICK-RELEASE ECONOMY 6537Q manufactured by Posey Products Llc.

Event Text Entries

[97947710] Product has been received but evaluation of the device was not completed at the time of this report. This report is based solely on the information provided by the customer. An investigation of similar events found that broken buckles were a result of wear and tear. Of those returned, the product was over 48 months old. There was no serious injury reported in any of the cases. Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device. The ifu states "always inspect before each use: check for broken stitches or parts; torn, cut or frayed material; or buckles that are cracked or broken and do not hold securely. Never use soiled or damaged products. " at this time, there is no evidence that a manufacturing non-conformity contributed to the reported complaint. Therefore, no corrective or preventative actions are necessary. All complaints are trended and reviewed by management on a monthly basis. As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted. Manufacturer reference file #(b)(4).
Patient Sequence No: 1, Text Type: N, H10


[97950091] Customer reported while transferring her father from the wheelchair to the bed, the buckle snapped and both the patient and caregiver fell. Neither suffered any injuries. Customer reported having the belt for a few years and did not report the date the incident occurred.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020362-2018-00001
MDR Report Key7208537
Date Received2018-01-19
Date of Report2018-03-13
Date Mfgr Received2018-03-13
Device Manufacturer Date2014-07-03
Date Added to Maude2018-01-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT FAMILY MEMBER OR FRIEND
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactWILLIAM HINCY
Manufacturer StreetPOSEY COMPANY 5635 PECK ROAD
Manufacturer CityARCADIA CA 91006
Manufacturer CountryUS
Manufacturer Postal91006
Manufacturer Phone6264433143
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTRANSFER BELT, QUICK-RELEASE ECONOMY
Generic NameAID, TRANSFER
Product CodeIKX
Date Received2018-01-19
Model Number6537Q
Catalog Number6537Q
Lot Number4184T107
OperatorPATIENT FAMILY MEMBER OR FRIEND
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPOSEY PRODUCTS LLC
Manufacturer Address5635 PECK RD ARCADIA CA 91006 US 91006


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-19

© 2026 FDA.report