Event Text Entries

[97976041] Transvaginal drain placement requested for pelvic abscess. Tvsono guided needle approach with sheathed needle performed. Difficult access due to height aka cranial level of collection and posterior anatomic location. First attempt with needle access failed as sheath popped out of the collection. Needle reinserted into sheath and access reattempted. Failure to gain access prompted exchange for the sheathed needle for a hard 18g trocar needle which was used to gain access and wire access allowed placement of transvaginal pelvic drain. Drain fell out on the inpatient floor the next day prompting of the pt and foreign body believed to be plastic piece of needle sheath identified.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5074692
MDR Report Key7209043
Date Received2018-01-19
Date of Report2018-01-17
Date of Event2017-12-04
Date Added to Maude2018-01-22
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOOK MEDICAL
Generic NameRING DRAINAGE CATHETER NEEDLESET
Product CodeDWO
Date Received2018-01-19
Model NumberREF G03388
Catalog NumberDLPN-40-25-RING
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerCOOK INCORPORATED
Manufacturer AddressBLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-01-19