ADAPTER, LL TO CATHETER TIP ST H93813601

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-01-22 for ADAPTER, LL TO CATHETER TIP ST H93813601 manufactured by Baxter Healthcare - Englewood.

Event Text Entries

[98080669] This event occurred in (b)(6) 2017. (b)(6). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[98080670] It was reported that there was a particle in the packaging of a catheter tip adapter. This occurred for thirty-eight (38) separate packages. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[104094977] Thirty-nine (39) actual samples were received for evaluation and an evaluation is complete. A visual inspection with magnification was performed and the reported condition was confirmed for three samples. One sample was found with foreign material inside of the sealed packaging. One sample had loose foreign material on the adapter surface which measured at 0. 10 square millimeters. The third sample was found with embedded particulate matter that measured at 0. 10 square millimeters on the adapter body. The reported condition was verified for these three samples. An assignable cause could not be determined. The reported issue was not verified for the other thirty-six samples. A batch review was conducted and there were no deviations found related to the reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2018-00309
MDR Report Key7210565
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-01-22
Date of Report2018-03-16
Date Mfgr Received2018-03-08
Date Added to Maude2018-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - ENGLEWOOD
Manufacturer Street14445 GRASSLANDS DR
Manufacturer CityENGLEWOOD CO 80112
Manufacturer CountryUS
Manufacturer Postal Code80112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADAPTER, LL TO CATHETER TIP ST
Generic NameDISPENSER, LIQUID MEDICATION
Product CodeKYX
Date Received2018-01-22
Returned To Mfg2018-01-24
Model NumberNA
Catalog NumberH93813601
Lot Number60078440
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - ENGLEWOOD
Manufacturer AddressENGLEWOOD CO


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-22

© 2024 FDA.report