RED -MLTI-DRG 6 PNL SCRN TST (DOA-164)

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2018-01-24 for RED -MLTI-DRG 6 PNL SCRN TST (DOA-164) manufactured by Alere San Diego, Inc..

Event Text Entries

[98296634] Investigation conclusion: the customer's observation was not replicated in-house with retention products. Retention devices were tested with in-house drug free donor urine and all devices showed expected negative results at read time for all analytes including coc, mamp, thc and bzo. No false positive results were observed during in-house testing. Manufacturing batch record review did not uncover any abnormalities. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[98296635] The customer reported receiving a false positive thc and mamp result on one donor using a multi-drug one step 6 drug screen test panel. The customer also reported receiving a false positive coc, mamp, thc, and bzo result on a second donor using another multi-drug one step 6 drug screen test panel. The false positive results occurred on 2 devices out of 100 devices used.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2018-00007
MDR Report Key7215917
Report SourceCONSUMER
Date Received2018-01-24
Date of Report2018-01-24
Date Mfgr Received2018-01-11
Date Added to Maude2018-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRED -MLTI-DRG 6 PNL SCRN TST (DOA-164)
Generic NameENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES
Product CodeDIO
Date Received2018-01-24
Model NumberDOA-164
Lot NumberDOA6090544
Device Expiration Date2018-08-31
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.