BLICKMAN MRI COMPATIBLE LINEN CART 7772 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2006-04-27 for BLICKMAN MRI COMPATIBLE LINEN CART 7772 * manufactured by Blickman, Inc..

Event Text Entries

[487872] Earlier this month, the blickman mri compatible laundry hamper was placed in close proximity to the mri scanner magnet and the laundry hamper was lifted and lodged into the scanner bore. No patients were involved in the incident and no one was hurt. A service call was placed and engineers (from the manufacturer of the mri scanner) came and dislodged the hamper from the magnet bore. A hand-held magnet was used to inspect the hamper, and it was found that two of the three vertical rods of the laundry hamper were magnetic. Product information states that all components are non-magnetic and mri compatible.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number721786
MDR Report Key721786
Date Received2006-04-27
Date of Report2006-04-27
Date of Event2006-04-25
Report Date2006-04-27
Date Reported to FDA2006-04-27
Date Added to Maude2006-06-05
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameBLICKMAN MRI COMPATIBLE LINEN CART
Generic NameHAMPER
Product CodeLDQ
Date Received2006-04-27
Model Number7772
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key710856
ManufacturerBLICKMAN, INC.
Manufacturer Address39 ROBINSON ROAD LODI NJ 07644 US


Patients

Patient NumberTreatmentOutcomeDate
10 2006-04-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.