MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2006-04-27 for BLICKMAN MRI COMPATIBLE LINEN CART 7772 * manufactured by Blickman, Inc..
[487872]
Earlier this month, the blickman mri compatible laundry hamper was placed in close proximity to the mri scanner magnet and the laundry hamper was lifted and lodged into the scanner bore. No patients were involved in the incident and no one was hurt. A service call was placed and engineers (from the manufacturer of the mri scanner) came and dislodged the hamper from the magnet bore. A hand-held magnet was used to inspect the hamper, and it was found that two of the three vertical rods of the laundry hamper were magnetic. Product information states that all components are non-magnetic and mri compatible.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 721786 |
MDR Report Key | 721786 |
Date Received | 2006-04-27 |
Date of Report | 2006-04-27 |
Date of Event | 2006-04-25 |
Report Date | 2006-04-27 |
Date Reported to FDA | 2006-04-27 |
Date Added to Maude | 2006-06-05 |
Event Key | 0 |
Report Source Code | User Facility report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | BLICKMAN MRI COMPATIBLE LINEN CART |
Generic Name | HAMPER |
Product Code | LDQ |
Date Received | 2006-04-27 |
Model Number | 7772 |
Catalog Number | * |
Lot Number | * |
ID Number | * |
Device Availability | Y |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 710856 |
Manufacturer | BLICKMAN, INC. |
Manufacturer Address | 39 ROBINSON ROAD LODI NJ 07644 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2006-04-27 |