GLUTEAL IMPLANT, ROUND, TEXTURED, SDM N/A CCB6-T-84

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-01-25 for GLUTEAL IMPLANT, ROUND, TEXTURED, SDM N/A CCB6-T-84 manufactured by Implantech Associates, Inc..

Event Text Entries

[98424068] Method: the actual device was returned and gross visual examination and examination under magnification was performed. A review of manufacturing records and quality control records was also performed. Results: code for "none" selected as no more appropriate code was located. A tear in the circumference of the device was observed. Conclusion: breakage or damage to medical devices on implantation is a known risk of the procedure. Product labeling warns that a prosthesis that appears damaged should not be implanted.
Patient Sequence No: 1, Text Type: N, H10


[98424069] Complainant reported that they received a gluteal implant that contained bubbles. They proceeded to insert the implant into a keller funnel, and it ripped. The problem occurred before there was any patient contact, and they had a backup available so there was no impact to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2028924-2018-00001
MDR Report Key7219175
Report SourceHEALTH PROFESSIONAL
Date Received2018-01-25
Date of Report2018-01-25
Date of Event2018-01-03
Date Mfgr Received2018-01-05
Device Manufacturer Date2017-10-10
Date Added to Maude2018-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CRAIG ARTHUR
Manufacturer Street6025 NICOLLE STREET, SUITE B
Manufacturer CityVENTURA CA 93003
Manufacturer CountryUS
Manufacturer Postal93003
Manufacturer Phone8053399415
Manufacturer G1IMPLANTECH ASSOCIATES, INC.
Manufacturer Street6025 NICOLLE STREET, SUITE B
Manufacturer CityVENTURA CA 93003
Manufacturer CountryUS
Manufacturer Postal Code93003
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGLUTEAL IMPLANT, ROUND, TEXTURED, SDM
Generic NameGLUTEAL IMPLANT
Product CodeMIB
Date Received2018-01-25
Returned To Mfg2018-01-10
Model NumberN/A
Catalog NumberCCB6-T-84
Lot Number873720
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerIMPLANTECH ASSOCIATES, INC.
Manufacturer Address6025 NICOLLE STREET, SUITE B VENTURA CA 93003 US 93003


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.