HARMONY VLED SURIGICAL LIGHTING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-25 for HARMONY VLED SURIGICAL LIGHTING SYSTEM manufactured by Steris Corporation - Montgomery.

Event Text Entries

[98577230] A steris technician arrived onsite to inspect the lighting system. The technician found that the plastic yoke cover had detached from the lighting system and that both screws were still intact with the plastic yoke cover assembly. The technician also observed cracks and abrasions on the surface of the lighting system. The harmony vled surgical lighting system operator manual states (pg. 1-4), "do not bump light heads into walls or other equipment. Use care when moving the arm upward to help avoid hitting the ceiling or objects above the arm. " the technician replaced the plastic cover for the subject vled surgical lighting system, confirmed the unit to be operating properly, and returned the unit to service. The reported event can be attributed to impact damage due to the lighthead bumping into other equipment or light heads. The lighting system was installed in 2011 and the unit has been in use for approximately 7 years. No additional issues have been noted.
Patient Sequence No: 1, Text Type: N, H10


[98577271] The user facility reported that the plastic yoke cover on their harmony vled surgical lighting system into the sterile field during a procedure. User facility personnel re-established the sterile field and the procedure was successfully completed without further incident. No report of injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1043572-2018-00010
MDR Report Key7220392
Date Received2018-01-25
Date of Report2018-01-25
Date of Event2017-12-29
Date Mfgr Received2017-12-29
Date Added to Maude2018-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927453
Manufacturer G1STERIS CORPORATION - MONTGOMERY
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY AL 36109
Manufacturer CountryUS
Manufacturer Postal Code36109
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHARMONY VLED SURIGICAL LIGHTING SYSTEM
Generic NameLIGHTING SYSTEM
Product CodeFST
Date Received2018-01-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - MONTGOMERY
Manufacturer Address2720 GUNTER PARK DRIVE EAST MONTGOMERY AL 36109 US 36109


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-25