[98847918] Only one of the three alarms summarized in this report was returned for evaluation. The returned unit does not sound when the pull magnet is taken off from the magnet plate due to a defective reed switch that likely failed due to wear and tear. The alarm was over 24 months. The other two alarms not returned were over 24 months old. The reported issue of the alarm not sounding cannot be confirmed. With the information available and without the return of the alarm, it cannot be determined what the alleged failure was related to. Based on previous complaint investigations and the age of the alarms, the likely cause of the event is wear / tear or impact such as dropping or bumping. At this time, there is no evidence that a manufacturing non-conformity contributed to the complaint. (b)(4)
Patient Sequence No: 1, Text Type: N, H10


[98847919] This report summarizes three malfunction events. The alarms have power but do not sound. (b)(4) asr report (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020362-2018-00008
MDR Report Key7223822
Date Received2018-01-27
Date of Report2018-01-23
Date Mfgr Received2018-01-23
Date Added to Maude2018-01-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactWILLIAM HINCY
Manufacturer StreetPOSEY COMPANY 5635 PECK ROAD
Manufacturer CityARCADIA CA 91006
Manufacturer CountryUS
Manufacturer Postal91006
Manufacturer Phone6264433143
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameKEEPSAFE ESSENTIAL
Generic NameFALL PREVENTION ALARM/SENSOR ATTACHED ONLY
Product CodePJP
Date Received2018-01-27
Model Number8373
Catalog Number8373
Lot NumberNA
Device Availability*
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPOSEY PRODUCTS LLC
Manufacturer AddressARCADIA CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-27