MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-30 for IMRIS ORT100 OPERATING ROOM TABLE ORT100 (T2) 108000-000 manufactured by Imris - Deerfield Imaging, Inc..
[98990209]
Service engineer observed that the hydraulic pressure regulator in the power pack was damaged which allowed system pressure to spike at the end of travel ranges. He replaced the hydraulic fluid filter, flushed the system, installed a new manifold, and ordered a new power pack. The table was tested to verify proper function. Internal capa issued.
Patient Sequence No: 1, Text Type: N, H10
[98990210]
Customer stated that the ort100 back section was lifting on it's own. This was observed during morning qa testing - no patient involvement.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3010326005-2018-00013 |
MDR Report Key | 7228814 |
Date Received | 2018-01-30 |
Date of Report | 2018-01-30 |
Date of Event | 2018-01-07 |
Date Mfgr Received | 2018-01-08 |
Device Manufacturer Date | 2009-09-23 |
Date Added to Maude | 2018-01-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR TODD SPERLING |
Manufacturer Street | 5101 SHADY OAK ROAD |
Manufacturer City | MINNETONKA MN 553434100 |
Manufacturer Country | US |
Manufacturer Postal | 553434100 |
Manufacturer Phone | 7632036344 |
Manufacturer G1 | IMRIS - DEERFIELD IMAGING, INC. |
Manufacturer Street | 5101 SHADY OAK ROAD |
Manufacturer City | MINNETONKA MN 553434100 |
Manufacturer Country | US |
Manufacturer Postal Code | 553434100 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | IMRIS ORT100 OPERATING ROOM TABLE |
Generic Name | OPERATING ROOM TABLE |
Product Code | KXJ |
Date Received | 2018-01-30 |
Model Number | ORT100 (T2) |
Catalog Number | 108000-000 |
Lot Number | N/A |
Operator | SERVICE PERSONNEL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | IMRIS - DEERFIELD IMAGING, INC. |
Manufacturer Address | 5101 SHADY OAK ROAD MINNETONKA MN 553434100 US 553434100 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-01-30 |