MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-01-30 for NEOBLUE MINI LED PHOTOTHERAPY SYSTEM 040823 manufactured by Natus Medical Incorporated.
        [98976657]
The complainant inquired about ordering information for a replacement led panel and constant current pcb, which natus technical service provided. Natus received a purchase order for a replacement constant current pcb on 1/5/2018, and the part was delivered to the complainant on 1/8/2018.
 Patient Sequence No: 1, Text Type: N, H10
        [98976658]
Natus received a report that the intensity of a neoblue mini device could not be adjusted to meet specifications. The complainant reported intensity of 2-3 uw/cm^2/nm when the potentiometer was adjusted to its maximum setting. A natus neoblue radiometer (calibration due 10/2018) was used to measure intensity. The neoblue mini service manual states that the light intensity specification is greater than or equal to 30 uw/cm^2/nm as measured by a neoblue radiometer. There was no report of death, serious injury, delay in treatment, or environmental/safety concerns.
 Patient Sequence No: 1, Text Type: D, B5
| Report Number | 3018859-2018-00080 | 
| MDR Report Key | 7229679 | 
| Report Source | USER FACILITY | 
| Date Received | 2018-01-30 | 
| Date of Report | 2018-01-02 | 
| Date Mfgr Received | 2017-01-02 | 
| Device Manufacturer Date | 2005-09-19 | 
| Date Added to Maude | 2018-01-30 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 3 | 
| Event Location | 3 | 
| Manufacturer Contact | KYLE HORTON | 
| Manufacturer Street | 5900 FIRST AVENUE SOUTH | 
| Manufacturer City | SEATTLE WA 98108 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 98108 | 
| Manufacturer Phone | 2068343921 | 
| Manufacturer G1 | NATUS MEDICAL INCORPORATED | 
| Manufacturer Street | 5900 FIRST AVENUE SOUTH | 
| Manufacturer City | SEATTLE WA 98108 | 
| Manufacturer Country | US | 
| Manufacturer Postal Code | 98108 | 
| Single Use | 3 | 
| Remedial Action | RL | 
| Previous Use Code | 3 | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | NEOBLUE MINI LED PHOTOTHERAPY SYSTEM | 
| Generic Name | NEOBLUE MINI | 
| Product Code | LBI | 
| Date Received | 2018-01-30 | 
| Model Number | 040823 | 
| Device Availability | * | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | N | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | NATUS MEDICAL INCORPORATED | 
| Manufacturer Address | 5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 1. Other | 2018-01-30 |