MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-01-30 for NEOBLUE MINI LED PHOTOTHERAPY SYSTEM 040823 manufactured by Natus Medical Incorporated.
[98976657]
The complainant inquired about ordering information for a replacement led panel and constant current pcb, which natus technical service provided. Natus received a purchase order for a replacement constant current pcb on 1/5/2018, and the part was delivered to the complainant on 1/8/2018.
Patient Sequence No: 1, Text Type: N, H10
[98976658]
Natus received a report that the intensity of a neoblue mini device could not be adjusted to meet specifications. The complainant reported intensity of 2-3 uw/cm^2/nm when the potentiometer was adjusted to its maximum setting. A natus neoblue radiometer (calibration due 10/2018) was used to measure intensity. The neoblue mini service manual states that the light intensity specification is greater than or equal to 30 uw/cm^2/nm as measured by a neoblue radiometer. There was no report of death, serious injury, delay in treatment, or environmental/safety concerns.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3018859-2018-00080 |
MDR Report Key | 7229679 |
Report Source | USER FACILITY |
Date Received | 2018-01-30 |
Date of Report | 2018-01-02 |
Date Mfgr Received | 2017-01-02 |
Device Manufacturer Date | 2005-09-19 |
Date Added to Maude | 2018-01-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | KYLE HORTON |
Manufacturer Street | 5900 FIRST AVENUE SOUTH |
Manufacturer City | SEATTLE WA 98108 |
Manufacturer Country | US |
Manufacturer Postal | 98108 |
Manufacturer Phone | 2068343921 |
Manufacturer G1 | NATUS MEDICAL INCORPORATED |
Manufacturer Street | 5900 FIRST AVENUE SOUTH |
Manufacturer City | SEATTLE WA 98108 |
Manufacturer Country | US |
Manufacturer Postal Code | 98108 |
Single Use | 3 |
Remedial Action | RL |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | NEOBLUE MINI LED PHOTOTHERAPY SYSTEM |
Generic Name | NEOBLUE MINI |
Product Code | LBI |
Date Received | 2018-01-30 |
Model Number | 040823 |
Device Availability | * |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | NATUS MEDICAL INCORPORATED |
Manufacturer Address | 5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2018-01-30 |