AQUABPLUS 2500 G02040108-US

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-02-01 for AQUABPLUS 2500 G02040108-US manufactured by Vivonic Gmbh.

Event Text Entries

[98984541] A manufacturer investigation is in progress. A supplemental medwatch report will be submitted upon completion of this activity.
Patient Sequence No: 1, Text Type: N, H10


[98984542] (b)(6), an installation logistics manager for vivonics, reported that an aquabplus 2500 reverse osmosis machine, located at a user facility, was removed from service due to a t1 test failure that occurred during set-up for hemodialysis (hd) use. It was reported to henry, by the user facility biomedical technician (biomed), that the sd card was burnt and there was an issue with the motherboard. This information was reported to fmcna rtg on (b)(6) 2018 during a conference call. The hd patients at this facility were able to receive treatment at another facility until the unit was repaired. There was no harm to any patients or individuals due to this event. There was no report of heat damage to the machine? S other components. The biomed replaced the sd card and motherboard to resolve the issue. The unit was returned to service at the user facility without a recurrence of the event as reported. The sd card was stated to be available to be returned to vivonic for physical evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010850471-2018-00001
MDR Report Key7234041
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-02-01
Date of Report2018-07-02
Date of Event2018-01-15
Date Mfgr Received2018-06-29
Date Added to Maude2018-02-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTHOMAS C. JOHNSON
Manufacturer Street920 WINTER ST.
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7816999499
Manufacturer G1VIVONIC GMBH
Manufacturer StreetKURUFUERST-EPPSTEIN-RING 4
Manufacturer CitySAILAUF 63877
Manufacturer CountryGM
Manufacturer Postal Code63877
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAQUABPLUS 2500
Generic NameSUBSYSTEM, WATER PURIFICATION
Product CodeFIP
Date Received2018-02-01
Model NumberAQUABPLUS 2500
Catalog NumberG02040108-US
Lot NumberN/A
Device AvailabilityY
Device AgeMO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVIVONIC GMBH
Manufacturer AddressKURUFUERST-EPPSTEIN-RING 4 SAILAUF 63877 GM 63877


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.