HELIOS 3000 TRACK MOUNTED DENTAL LIGHT HL3T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-02 for HELIOS 3000 TRACK MOUNTED DENTAL LIGHT HL3T manufactured by Pelton & Crane.

Event Text Entries

[99194215] Upon evaluation by the distributor it was determined the roll pins and set screws were not installed by the distributor during installation. The roll pins and set screws will prevent the light from unscrewing from the trolley after installation. The pelton & crane installation instructions clearly states to properly install the roll pins and set screws during installation of the track light. The installation instructions also lists warnings to ensure the roll pins and set screws are installed. The distributor service technician ordered the roll pins and set screws and will properly install the light.
Patient Sequence No: 1, Text Type: N, H10


[99194216] It was reported that a dentist had finished treating a patient when the pelton & crane helios dental light started to fall down towards the ground. The dr. Caught the light with her hand in the air and layed it down. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2018-00003
MDR Report Key7235771
Date Received2018-02-02
Date of Report2018-01-11
Date of Event2018-01-11
Date Mfgr Received2018-01-11
Device Manufacturer Date2009-11-19
Date Added to Maude2018-02-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1PELTON & CRANE
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHELIOS 3000 TRACK MOUNTED DENTAL LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2018-02-02
Model NumberHL3T
Catalog NumberHL3T
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-02