MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-03 for 1251QCS manufactured by Bard Brachytherapy, Inc. -1424526.
[99102410]
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard. The device was not returned.
Patient Sequence No: 1, Text Type: N, H10
[99102411]
It was reported by the complainant that the cartridge was not dispensing the seeds properly. It is unknown if the patient received all of the seeds. It was impossible to set up the i125 seeds in the applicator quicklink. The applicator looked as if there was no more seeds in the cartridge. The procedure was completed putting the seeds in the applicator by hand, which could be a radiation protection problem. The last cartridge was used, as it was unable to be changed. It is unknown if the patient received all of the seeds.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1018233-2018-00320 |
MDR Report Key | 7239959 |
Date Received | 2018-02-03 |
Date of Report | 2018-02-16 |
Date Mfgr Received | 2018-02-16 |
Date Added to Maude | 2018-02-03 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | AMY GRAVLEY |
Manufacturer Street | 8195 INDUSTRIAL BLVD |
Manufacturer City | COVINGTON GA 30014 |
Manufacturer Country | US |
Manufacturer Postal | 30014 |
Manufacturer Phone | 7707846100 |
Manufacturer G1 | BARD BRACHYTHERAPY, INC. -1424526 |
Manufacturer Street | 295 EAST LIES ROAD |
Manufacturer City | CAROL STREAM IL 60188 |
Manufacturer Country | US |
Manufacturer Postal Code | 60188 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Generic Name | QUICKLINK |
Product Code | KXK |
Date Received | 2018-02-03 |
Catalog Number | 1251QCS |
Lot Number | BBBZ0022 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BARD BRACHYTHERAPY, INC. -1424526 |
Manufacturer Address | 295 EAST LIES ROAD CAROL STREAM IL 60188 US 60188 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-02-03 |