COC THC MOP AMP MAMP BZO BAR MTD MDMA OX DOA-1127

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional, report with the FDA on 2018-02-06 for COC THC MOP AMP MAMP BZO BAR MTD MDMA OX DOA-1127 manufactured by Alere San Diego, Inc..

Event Text Entries

[99789666] Investigation conclusion: the customer's observation was not replicated in-house with retention and return products. Retention and return devices were tested with in-house drug free donor urine and all devices showed expected negative results at read time for all analytes including amp, mamp, mop, coc, and thc. No false positive results were observed during in-house testing. Manufacturing batch record review did not uncover any abnormalities. Based on the information available, there is no indication of a product deficiency. A capa, capa-(b)(4), has been initiated to address the late documentation of this case.
Patient Sequence No: 1, Text Type: N, H10


[99789667] The customer reported receiving positive results using the multi-drug 12 panel key cup for multiple analytes as compared to negative results received at the laboratory between september 2017 and december 2017. The customer reported the issue was isolated to a single patient. Although the customer was unable to confirm the specific issue observed, pictures were submitted showing positive results for amp, mamp, coc, mop, and thc using the multi-drug 12 panel key cup. Although requested, no additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2018-00017
MDR Report Key7245041
Report SourceCONSUMER,HEALTH PROFESSIONAL,
Date Received2018-02-06
Date of Report2018-02-06
Date Mfgr Received2017-12-12
Date Added to Maude2018-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOC THC MOP AMP MAMP BZO BAR MTD MDMA OX
Generic NameENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES
Product CodeDIO
Date Received2018-02-06
Returned To Mfg2018-01-19
Model NumberDOA-1127
Lot NumberDOA7050432
Device Expiration Date2019-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-06

© 2026 FDA.report