SITUATE 01-0031

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-06 for SITUATE 01-0031 manufactured by Mitg - Rf Surgical Systems.

Event Text Entries

[99781804] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[99781805] According to the reporter, the device would not detect the sponges. No patient involved. No further information available.
Patient Sequence No: 1, Text Type: D, B5


[105905859] Evaluation summary: one device was received for evaluation. This device had not been used in the treatment or diagnosis of the patient. The returned product met specification as received. Visual inspection found no defects. The reported condition was not confirmed. The investigation found all the coils passed testing on the tester. The mat was plugged into the console detected sponges. The investigation found the device to function normally and within specifications. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005883396-2018-00014
MDR Report Key7245776
Date Received2018-02-06
Date of Report2018-03-02
Date Mfgr Received2018-02-14
Date Added to Maude2018-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSHARON MURPHY
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal80301
Manufacturer Phone2034925267
Manufacturer G1MITG - RF SURGICAL SYSTEMS
Manufacturer Street2101 FARADAY
Manufacturer CityCARLSBAD CA 92008
Manufacturer CountryUS
Manufacturer Postal Code92008
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSITUATE
Generic NameERYTHROCYTE SUSPENSION, MULTI SPECIES, SEROLOGICAL REAGENT AND EQUIPMENT
Product CodeLHW
Date Received2018-02-06
Returned To Mfg2018-01-31
Model Number01-0031
Catalog Number01-0031
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMITG - RF SURGICAL SYSTEMS
Manufacturer Address2101 FARADAY CARLSBAD CA 92008 US 92008


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-06