INVACARE RELIANT 450 BATTERY-POWERED LIFT RPL450-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-07 for INVACARE RELIANT 450 BATTERY-POWERED LIFT RPL450-1 manufactured by Invacare Taylor Street.

Event Text Entries

[99330921] The dealer provided a photograph of the electric actuator. The photograph shows that the mounting holes on the shaft extension of the actuator are worn through, which would result in the actuator becoming detached from the boom at the mounting bracket, as described. The lift was returned to invacare for an evaluation, and it was identified that the bushing, as well as the finger guard, were missing from the actuator assembly. This likely caused excessive play in the actuator end, that over time caused wear and lead to the actuator end breaking. It has been determined that the cause of the failure in aging devices (outside of their 8-year life expectancy, as is the case for this lift) is likely due to improper maintenance and/or placement of the bushing during initial assembly/maintenance by the purchaser of the lift. If the bushing is not in place then the metal-to-metal contact, between the shaft extension of the actuator and the shoulder bolt, will cause wear after considerable use. Review of current packaging confirmed that the bushing is assembled onto the shoulder bolt/nut on the lift boom when shipped from invacare's manufacturing site. The maintenance safety inspection checklist within the lift owner's manual instructs to inspect the hardware connecting the actuator assembly to the mast and boom and to check for wear or deterioration on a monthly basis. Any defective parts should be replaced immediately.
Patient Sequence No: 1, Text Type: N, H10


[99330922] The dealer reported that her customer alleged that two certified nursing assistants were transferring the patient from a bed to a chair, and the rpl450-1 lift failed causing the patient to fall. The patient was sore following the incident but did not require any medical intervention. The facility indicated that the failure occurred where the actuator attaches to the boom of the lift; the bolt fell out and now the boom will not hold in a vertical position.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1525712-2018-00013
MDR Report Key7248218
Date Received2018-02-07
Date of Report2018-01-08
Date Mfgr Received2018-01-08
Device Manufacturer Date2005-01-01
Date Added to Maude2018-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JASON FIEST
Manufacturer StreetONE INVACARE WAY
Manufacturer CityELYRIA OH 44035
Manufacturer CountryUS
Manufacturer Postal44035
Manufacturer Phone8003336900
Manufacturer G1INVACARE TAYLOR STREET
Manufacturer Street1200 TAYLOR STREET
Manufacturer CityELYRIA 44035
Manufacturer CountryUS
Manufacturer Postal Code44035
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameINVACARE RELIANT 450 BATTERY-POWERED LIFT
Generic NameLIFT, PATIENT, AC-POWERED
Product CodeFNG
Date Received2018-02-07
Returned To Mfg2018-01-24
Model NumberRPL450-1
Catalog NumberRPL450-1
OperatorNURSING ASSISTANT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINVACARE TAYLOR STREET
Manufacturer Address1200 TAYLOR STREET ELYRIA OH 44035 US 44035