N
Patient 1
THE CUSTOMER RETURNED FIVE OF SIX ELECTRODES TO OLYMPUS FOR EVALUATION. A VISUAL INSPECTION OF THE THIRD ELECTRODE (A22251C LOT# P1780004) FOUND THE WIRE THAT HOLDS THE ROLLER BALL AT THE DISTAL TIP IS BROKEN AND SEPARATED FROM ONE OF THE SILVER COLORED RESECTION POSTS. THE ROLLER BALL IS INTACT. A MICROSCOPE INSPECTION OF THE BREAKAGE POINT FOUND THE ROLLER BALL AND THE RESECTION POST ARE DISCOLORED. THE INSULATION AT THE DISTAL TIP ON BOTH RESECTION POSTS ARE CHARRED, BURNED AND MELTED. NO PART OF WIRE THAT HOLDS THE ROLLER IN PLACE IS SUSPECTED TO BE MISSING. THE CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AS THE USER FACILITY DID NOT RETURN THE CONCOMITANT DEVICES FOR TESTING. HOWEVER, BASED ON SIMILAR REPORTED EVENTS THE MOST PROBABLE CAUSE OF THE REPORTED EVENT COULD BE ATTRIBUTED TO THE ELECTRODE COMING INTO CONTACT (UNINTENDED) WITH OTHER METAL PARTS, E.G. SURGICAL INSTRUMENTS WHILE THE HIGH-FREQUENCY OUTPUT WAS ACTIVATED AND HIGHER OUTPUT SETTINGS ON THE ELECTROSURGICAL GENERATOR COULD CAUSE THE ELECTRODE TO BECOME DAMAGED.