GORE? VIATORR? TIPS ENDOPROSTHESIS PTB8108275

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-02-08 for GORE? VIATORR? TIPS ENDOPROSTHESIS PTB8108275 manufactured by W.l. Gore & Associates.

Event Text Entries

[99364131] A review of the manufacturing records for the device verified that the lot met all pre-release specifications. The device remains implanted so no engineering investigation could be performed. (b)(4). An issue in which a medical device has an obstructed lumen.
Patient Sequence No: 1, Text Type: N, H10


[99364132] It was reported that the physician was implanted a gore? Viatorr? Tips endoprosthesis for a transjugular intrahepatic portosystemic shunt (tips) procedure. The device was implanted on (b)(6) 2017. On (b)(6) 2017 the device was imaged and stent occlusion was observed. The device had initially been dilated to 8mm in diameter. The physician went back and re-dilated the device to 10mm. Following the procedure the stent was patent and the patient was doing well.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007284313-2018-00049
MDR Report Key7252084
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-02-08
Date of Report2018-01-17
Date of Event2017-11-30
Device Manufacturer Date2017-07-17
Date Added to Maude2018-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactASHLEY MAROSTICA
Manufacturer Street1500 N. 4TH STREET
Manufacturer CityFLAGSTAFF AZ
Manufacturer Phone9285263030
Manufacturer G1MEDICAL PHOENIX 1 B/P
Manufacturer Street32360 N. NORTH VALLEY PARKWAY
Manufacturer CityPHOENIX AZ 85085
Manufacturer CountryUS
Manufacturer Postal Code85085
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGORE? VIATORR? TIPS ENDOPROSTHESIS
Generic NameSHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS
Product CodeMIR
Date Received2018-02-08
Catalog NumberPTB8108275
Lot Number16507194
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerW.L. GORE & ASSOCIATES
Manufacturer AddressFLAGSTAFF AZ


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-02-08

© 2026 FDA.report