MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-08 for DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS 8065751512 manufactured by Alcon Research, Ltd. - Alcon Precision Device.
[99534830]
Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[99534831]
A surgeon reported that the patient's posterior capsule was found to be damaged after using the i/a tip with a flaw during a cataract with intraocular lens implant procedure. Additional information has been requested.
Patient Sequence No: 1, Text Type: D, B5
[109977086]
One polymer curved i/a tip was returned for evaluation for the report of a burr or flaw that may have resulted in posterior capsule damage. No lot number was identified with this complaint; therefore, a lot history review could not be conducted. A visual inspection was performed and deemed conforming. No flaw, burr, or other visual nonconformance was present with the i/a tip. The complaint evaluation does not confirm the polymer ia tip has nonconforming burr or a flaw. The tip was manufactured to specification. The reason for the posterior capsule damage cannot be determined from this evaluation. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2523835-2018-00061 |
MDR Report Key | 7253168 |
Date Received | 2018-02-08 |
Date of Report | 2018-04-25 |
Date of Event | 2018-01-15 |
Date Mfgr Received | 2018-04-13 |
Date Added to Maude | 2018-02-08 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. BRYAN BLAKE |
Manufacturer Street | 6201 SOUTH FREEWAY MAIL STOP AB2-6 |
Manufacturer City | FORT WORTH TX 76134 |
Manufacturer Country | US |
Manufacturer Postal | 76134 |
Manufacturer Phone | 8176152230 |
Manufacturer G1 | ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE |
Manufacturer Street | 714 COLUMBIA AVENUE |
Manufacturer City | SINKING SPRING PA 19608 |
Manufacturer Country | US |
Manufacturer Postal Code | 19608 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 0 |
Brand Name | DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS |
Generic Name | DEVICE, IRRIGATION, OCULAR SURGERY |
Product Code | KYG |
Date Received | 2018-02-08 |
Returned To Mfg | 2018-02-09 |
Model Number | NA |
Catalog Number | 8065751512 |
Lot Number | ASKU |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE |
Manufacturer Address | 714 COLUMBIA AVENUE SINKING SPRING PA 19608 US 19608 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2018-02-08 |