DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS 8065751512

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-08 for DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS 8065751512 manufactured by Alcon Research, Ltd. - Alcon Precision Device.

Event Text Entries

[99534830] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[99534831] A surgeon reported that the patient's posterior capsule was found to be damaged after using the i/a tip with a flaw during a cataract with intraocular lens implant procedure. Additional information has been requested.
Patient Sequence No: 1, Text Type: D, B5


[109977086] One polymer curved i/a tip was returned for evaluation for the report of a burr or flaw that may have resulted in posterior capsule damage. No lot number was identified with this complaint; therefore, a lot history review could not be conducted. A visual inspection was performed and deemed conforming. No flaw, burr, or other visual nonconformance was present with the i/a tip. The complaint evaluation does not confirm the polymer ia tip has nonconforming burr or a flaw. The tip was manufactured to specification. The reason for the posterior capsule damage cannot be determined from this evaluation. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2523835-2018-00061
MDR Report Key7253168
Date Received2018-02-08
Date of Report2018-04-25
Date of Event2018-01-15
Date Mfgr Received2018-04-13
Date Added to Maude2018-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRYAN BLAKE
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8176152230
Manufacturer G1ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
Manufacturer Street714 COLUMBIA AVENUE
Manufacturer CitySINKING SPRING PA 19608
Manufacturer CountryUS
Manufacturer Postal Code19608
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameDISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS
Generic NameDEVICE, IRRIGATION, OCULAR SURGERY
Product CodeKYG
Date Received2018-02-08
Returned To Mfg2018-02-09
Model NumberNA
Catalog Number8065751512
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
Manufacturer Address714 COLUMBIA AVENUE SINKING SPRING PA 19608 US 19608


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-02-08