CHROMID? VRE AGAR 43004

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-02-09 for CHROMID? VRE AGAR 43004 manufactured by Biom?rieux Sa.

Event Text Entries

[100124408] A customer from (b)(6) reported a false resistant result for a vancomycin-susceptible enterococcus faecalis strain from a rectal swab, in association with the chromid? Vre agar. The customer stated the vancomycin-susceptible strain with typical color of vre, exhibited growth on the chromid? Vre agar after 24 hours; organism growth should have been inhibited. The strain was tested with the vitek? 2 ast-p586 card and was vancomycin susceptible (mic=1). The pcr result was van-a/b negative. The customer was not able to retest, as the isolate could not be re-cultivated on a vre plate. The customer reported the initial incorrect result was reported to a physician but was corrected. The customer stated there was no impact to patient treatment. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00014
MDR Report Key7260428
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-02-09
Date of Report2018-04-12
Date Mfgr Received2018-03-20
Device Manufacturer Date2017-12-06
Date Added to Maude2018-02-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, FR 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? VRE AGAR
Generic NameCHROMID? VRE AGAR
Product CodeJSO
Date Received2018-02-09
Catalog Number43004
Lot Number1006145660
ID Number03573026257972
Device Expiration Date2018-03-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-09

© 2026 FDA.report