KONICA MINOLTA AIR-SHIELDS JAUNDICE METER JM-103 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2006-06-06 for KONICA MINOLTA AIR-SHIELDS JAUNDICE METER JM-103 * manufactured by Draeger Medical, Inc..

Event Text Entries

[486603] Nurse took a reading with the jaundice meter and got a reading of 7. 6. Serum that was drawn and sent at the same time was 12. 7, with a difference of 5. 1 points on baby's bilirubin level. All four jaundice meters were pulled because the department wasn't sure which unit had been used. Calibration on all four meters was done. They all calibrated out ok. All four units are being sent back to the manufacturer for testing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number726097
MDR Report Key726097
Date Received2006-06-06
Date of Report2006-06-06
Date of Event2006-05-30
Report Date2006-06-06
Date Reported to FDA2006-06-06
Date Added to Maude2006-06-16
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameKONICA MINOLTA AIR-SHIELDS JAUNDICE METER
Generic NameJAUNDICE METER
Product CodeLMX
Date Received2006-06-06
Returned To Mfg2006-06-06
Model NumberJM-103
Catalog Number*
Lot Number*
ID Number*
OperatorNURSE
Device AvailabilityR
Device Age1 YR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key714345
ManufacturerDRAEGER MEDICAL, INC.
Manufacturer Address3135 QUARRY ROAD TELFORD PA 18969 US

Device Sequence Number: 2

Brand NameKONICA MINOLTA AIR-SHIELDS JAUNDICE METER
Generic NameJAUNDICE METER
Product CodeLMX
Date Received2006-06-06
Returned To Mfg2006-06-06
Model NumberJM-103
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityR
Device Age1 YR
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key714346
ManufacturerDRAEGER MEDICAL, INC.
Manufacturer Address3135 QUARRY ROAD TELFORD PA 18969 US

Device Sequence Number: 3

Brand NameKONICA MINOLTA AIR-SHIELDS JAUNDICE METER
Generic NameJAUNDICE METER
Product CodeLMX
Date Received2006-06-06
Returned To Mfg2006-06-06
Model NumberJM-103
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityR
Device Age1 YR
Implant FlagN
Date Removed*
Device Sequence No3
Device Event Key714347
ManufacturerDRAEGER MEDICAL, INC.
Manufacturer Address3135 QUARRY ROAD TELFORD PA 18969 US

Device Sequence Number: 4

Brand NameKONICA MINOLTA AIR-SHIELDS JAUNDICE METER
Generic NameJAUNDICE METER
Product CodeLMX
Date Received2006-06-06
Returned To Mfg2006-06-06
Model NumberJM-103
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityR
Device Age1 YR
Implant FlagN
Date Removed*
Device Sequence No4
Device Event Key714348
ManufacturerDRAEGER MEDICAL, INC.
Manufacturer Address3135 QUARRY ROAD TELFORD PA 18969 US


Patients

Patient NumberTreatmentOutcomeDate
10 2006-06-06

© 2025 FDA.report