ELECSYS BRAHMS PCT E801 07301715190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-02-12 for ELECSYS BRAHMS PCT E801 07301715190 manufactured by Roche Diagnostics.

Event Text Entries

[100260026] (b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[100260027] The customer received questionable high pct brahms (roche) elecsys e2g (procalcitonin) results during correlation testing for a new cobas 8000 e 801 module. The samples were initially tested (b)(6) 2018 on a cobas 8000 e 801 module and the results were reported outside of the laboratory. On (b)(6) 2018, the samples were repeated on a cobas e 411 immunoassay analyzer and two cobas 8000 e 801 modules. The repeat results from the cobas e 411 immunoassay analyzer were believed to be correct. There was no allegation of an adverse event. The reagent lot number and expiration date were requested but were not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-00399
MDR Report Key7263126
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-02-12
Date of Report2018-03-19
Date of Event2018-01-04
Date Mfgr Received2018-01-25
Date Added to Maude2018-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberRES 79462
Event Type3
Type of Report3

Device Details

Brand NameELECSYS BRAHMS PCT
Generic NameDEVICE TO DETECT AND MEASURE PROCALCITONIN (PCT) IN HUMAN CLINICAL SPECIMENS
Product CodePMT
Date Received2018-02-12
Model NumberE801
Catalog Number07301715190
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.