HF-RESECTION ELECTRODE, ROLLER, 24-28 FR A22251C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2018-02-12 for HF-RESECTION ELECTRODE, ROLLER, 24-28 FR A22251C manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[100041932] The customer returned five of six electrodes to olympus for evalauation. A visual inspection of the fourth electrode (a22251c lot# p1780004) found the roller ball and wire at the distal tip of the electrode are broken and missing. A microscope inspection of the breakage points found the insulation on both resection posts are heavily charred, burned and melted at the distal tip. The cause of the reported event could not be conclusively determined as the user facility did not return the concomitant devices for testing. However, based on similar reported events the most probable cause of the reported event could be attributed to the electrode coming into contact (unintended) with other metal parts, e. G. Surgical instruments while the high-frequency output was activated and higher output settings on the electrosurgical generator could cause the electrode to become damaged.
Patient Sequence No: 1, Text Type: N, H10


[100041933] Olympus was informed that during a transurethral resection of prostate (turp) procedure, the loop wire on two (a22202c) electrodes and the roller of four (a22251c) electrodes broke off inside of a patient. The device fragments were retrieved from the patient. The procedure was completed using an unspecified electrode. No patient injury was reported. Five of 6.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2018-00099
MDR Report Key7264826
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2018-02-12
Date of Report2018-03-01
Date of Event2018-01-12
Date Mfgr Received2018-02-14
Device Manufacturer Date2017-08-01
Date Added to Maude2018-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CONNIE TUBERA
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone408935-512
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHF-RESECTION ELECTRODE, ROLLER, 24-28 FR
Product CodeGCP
Date Received2018-02-12
Returned To Mfg2018-01-30
Model NumberA22251C
Catalog NumberA22251C
Lot NumberP1780004
ID Number14042761036815
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG, HAMBURG US


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.