STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-02-13 for STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[100148311] Patient's exact age is unknown; however it was reported that the patient was over the age of 18. (b)(4). A visual analysis of the returned device, found that it was returned- approximately half open/closed. The blue sheath was pulled back to expose the coil. The coil was in its intended shape. The blue/green heat shrink is damaged near the distal end of the coil. As a result of the damages, the device opens freely but will not close completely. Since the damage occurred during unpacking, the most probable root cause is "handling damage". The complaint was caused by handling of the device or portion of the device without direct patient contact either during unpacking, preparation, or shipping; at the end of the procedure; or when packaging for return. The device history record (dhr) review found the device met all manufacturing specifications.
Patient Sequence No: 1, Text Type: N, H10


[100148312] It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was used in the ureter during a ureteroscopy (urs) procedure performed on (b)(6) 2017. According to the complainant, during unpacking, the stone cone nitinol urological retrieval coil was opened and it was found that the coil was not in its regular shape. The procedure was completed with another stone cone nitinol urological retrieval coil. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be? Stable?. This event has been deemed an mdr-reportable event based on investigation results which revealed that the coil/cone coil coating peeled/shared/frayed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-00259
MDR Report Key7267631
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-02-13
Date of Report2018-01-25
Date of Event2017-10-26
Date Mfgr Received2018-01-25
Device Manufacturer Date2017-04-03
Date Added to Maude2018-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2018-02-13
Returned To Mfg2018-01-02
Model NumberM0063903200
Catalog Number390-320
Lot NumberG35698
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752